Description
BRT LB06 Process calibrator can be used as a high precision mV mA calibrator, 4-20mA loop calibrator, temperature calibrator, frequency calibrator, Hart operator, Modbus Master/Slave meter, digital multimeter, etc. It is an upgrade premium version of LB02 process calibrator. It covers all the functions of LB02 process calibrator and add frequency, Hart, Modbus function, etc. Its accuracy is up to 0.025%, its button has corresponding function description, user can use it without difficulties. Inside the LB06 process calibrator, there are high precision AD, DA chips, highly efficient circuits used to ensure its high accuracy and stable performance. This multifunction process calibrator is an ideal low-cost high precision process controlling signal calibrator with multimeter.
Main Features
● High Precision 0.025%+1/2 digits, high reliability & stability in signal output and measurement.
● User friendly design, pocket size: 122 x 80 x 27mm, put it into your pocket, save energy, save cost.
● Output simulation: 0/4-20mA source/passive, 0-11V, 0-110mV, 20-400Ω, R, E, S, K, J, B, T, N type thermocouple, PT1000, PT100, CU50, 24V outputs, frequency, etc.
● Measurement: -30-+30V, -150- +150mV, -30mA to +30mA, 0-999.9Ω, R, E, S, K, J, B, T, N type thermocouple, PT100, Pt1000 CU50, 24V, 0-24mA, Ohm, frequency, etc.
● Signal generator, multifunction process calibrator and digital multimeter all-in-one.
● Have standard MODBUS Master, MODBUS Slave, Hart communicator or operator function.
● Have description in V, mV, mA, RTD, TC, Step, Ω, Hz, MODBUS, Hart keys, use it easily.
● Have automatic output mode, manual output mode, customized output mode, signal conversion, curves output, full function operation, etc.
● High resolution color LCD display with back light for different lighting conditions, large screen clear display.
● Fast response, no delay in signal source simulation and signal testing or measurement.
● Have battery capacity bar to check the battery status easily.
● Precision AD, DA, reference voltage, OP circuits are used to ensure high accuracy, high performance and high stability.
● Have 4 units of PPTC and a quick-break fuse to protect the instrument from damaged by high voltage.
Output Parameters
FUNC | OUT UNIT | RANGE | SET STEP | ACCURACY (SP: setting point) |
DC. V | V | 0~11.00V | 0.01/0.1/1 | ±0.025%SP ±0.004V |
DC. mV | mV | 0~110.00mV | 0.1/1/10 | ±0.025%SP ±0.03mV |
DC. mA | Sink/Source/Program | 0~24.00mA | 0.01/0.1/1/4 | ±0.025%SP ±0.004mA |
Ohm | Ω | 0~400Ω | 1/10/100 | ±0.025%SP ±0.3Ω |
400~2200Ω | 1/10/100 | ±0.05%SP ±1.5Ω | ||
Frequency | KHz | 0-10 | 0.01/0.1/1 | ±0.05%SP ±2Hz |
TC | R | 0~1700℃ | 1/10/100 | ±0.1%SP ±3℃ |
S | 0~1600℃ | |||
B | 500~1800℃ | |||
K | -200~1370℃ | ±0.05%SP ±0.5℃(K) ±0.05%SP ±0.3℃(E) ±0.05%SP ±0.4℃(J) ±0.05%SP ±0.5℃(T) ±0.05%SP ±0.8℃(N) | ||
E | -200~1000℃ | |||
J | -200~1200℃ | |||
T | -200~400℃ | |||
N | -200~1300℃ | |||
RTD | Pt1000 | -50~300℃ | 1/10/100 | ±0.1%SP±0.5℃ |
Pt100 | -200~850℃ | ±0.1%SP ±0.8℃ | ||
Cu50 | -50~150℃ | ±0.1%SP±1.5℃ | ||
24V | 24V mA | 24V: Cannot be set. Current measurement:0~24.000 mA | NONE |
Measurement Parameters
FUNC | UNIT | RANGE | RESOLUTION | ACCURACY (RD: Read number) |
V | V | -30~30.000V | 0.001V | ±0.025%RD ±0.004V |
DC. mV | mV | -150~150.00mV | 0.01mV | ±0.025%RD ±0.02mV |
DC. mA | mA | -30~30.000mA | 0.001mA | ±0.025%RD ±0.004mA |
Ohm | Ω | 0~999.9Ω | 0.1Ω | ±0.025%RD ±0.2Ω |
1000-2500 | 1Ω | ±0.025%RD ±1Ω | ||
Frequency | KHz | 0-10 | 0.001 | ±0.05%RD ±2Hz |
TC | R | 0~1700℃ | 0.1℃ | ±0.1%RD ±3℃ |
S | 0~1600℃ | |||
B | 500~1800℃ | |||
K | -200~1370℃ |
±0.025%RD ±0.5℃(K) ±0.025%RD ±0.3℃(E) ±0.025%RD ±0.4℃(J) ±0.05%RD ±0.5℃(T) ±0.025%RD ±0.8℃(N) | ||
E | -200~1000℃ | |||
J | -200~1200℃ | |||
T | -200~400℃ | |||
N | -200~1300℃ | |||
RTD | Pt1000 | -50~300℃ | 0.1℃ | ±0.1%RD ±0.5℃ |
Pt100 | -200~850℃ | ±0.1%RD ±0.8℃ | ||
Cu50 | -50~150℃ | ±0.1%RD ±1.5℃ | ||
Range switch | mA | 0~9990 engineering unit (for current measurement and measure the current of 24V output) | Max. 4, Min. 1 (related to engineering quantity.) |
* Ohm, RTD output & measurement: Lead resistance error is not included in the accuracy.
* Thermocouple output & measurement: Cold junction error is not included in the accuracy.
Testing Show
*How to Order it:
•A. Visa, Master, Credit card pay online directly by clicking [Add to Cart] button above and checkout accordingly (Recommended).
•B. Paypal pay online secured if you have a Paypal account, or Credit card payment processed by Paypal online directly by clicking [Add to Cart] button above.
•C. Retailer order or Bulk order via Bank wire transfer, please contact us to get more details.
Reviews
There are no reviews yet.